Heat antigen retrieval
Material:
Hot plate
2000ml glass beaker
250ml plastic container
slides rack
slides container
Thermometer
Buffer: Bulls Eye (BULL1000L2JX BioCare) or 0.01M sodium citrate buffer, pH 6.0
Procedure:
- Fill glass beaker with 1000ml ddH2O.
- Put a thermometer in the beaker.
- Put beaker on hot plate and turn on the hotplate.
- Check water temperature until it is stable at the desire temperature (95-99°C or 102-104°C).
- Use a 250ml glass bottle and fill with 190ml ddH2O. Warm 30 minutes before use it.
- Deparaffinize the slides and keep slides in the container with ddH2O.
- Take bottle from water bath and add 10ml 20X bulls eye solution to the bottle. Shack well.
- Empty ddH2O in slides container and pour 1X bulls eye solution to slides container.
- Put Container back to water bath as soon as possible.
- Heat for 20 min.
- Wash slide with ddH2O 2-3 times Slides are ready for immunostaining.